Specifications

Dimensions:
- 19” standard enclosure, 4HE
- 45 x 18 x 55 cm (w*h*d)
- footprint (w*d):
45*55 (landscape position)
18*55 (portrait position)
- weight: 22 kg (2 ch. Instrument)

Power:
100-240V DC, 47-63 Hz,
12A max

Heated zones:
up to 10

Electronic Gas Control
- connection: 1/8” Swagelock
- max 7 modules;
up to 14 channels
- max inlet pressure 1000kPa
- max outlet pressure
350/700 kPa
- max splitflow: 100 ml/min
- carrier gas He, Nr, Ar, (H2)
- detector gas depending on
detector

Analysis channels:
1-4; extendable
Valve oven:
- independently heated, 40-150 °C
- up to 3 valves
- gas sampling or liquid sampling valves
- diaphragm valves with internal purge

Column ovens:
- independently heated, 40-250 °C
- up to 3

Detectors:
TCD, FID, PDD, PED, PID, PFPD, MS
Columns
- capillary 0,32/0,25mm ID and smaller
- wide bore MXT 0.53mm ID
- micropacked 1/16” OD

Column switching
- diaphragm valve
- Deans backflush or heartcut

Sampling
- inlet temperature: max. 150 °C,
- inlet pressure: max. 21 bar
(depending on used valve)
- connection: 1/8” Swagelock
- injection volume: 0.06 ul to
5ml (loop)

Repeatability
- 0.2% RSD or better (n=10,
Ethane, TCD detection)

Data communications
- LAN for GC parameter control and data transfer
(EZChrom/ChromQuest/OpenLab); GC editor available
- 8 * external events 24V (valve control)
- 4 * SSR output
- 24 VDC
- 3 * digital input (remote start-stop-prep)
- optional analogue output (0-1V) for detector signals

Environmental conditions
- Ambient operating temperature: 5-45 °C
- Ambient operating humidity: 5-95 %,
(non-condensable)