Dimensions: - 19” standard enclosure, 4HE - 45 x 18 x 55 cm (w*h*d) - footprint (w*d): 45*55 (landscape position) 18*55 (portrait position) - weight: 22 kg (2 ch. Instrument) Power: 100-240V DC, 47-63 Hz, 12A max Heated zones: up to 10 Electronic Gas Control - connection: 1/8” Swagelock - max 7 modules; up to 14 channels - max inlet pressure 1000kPa - max outlet pressure 350/700 kPa - max splitflow: 100 ml/min - carrier gas He, Nr, Ar, (H2) - detector gas depending on detector Analysis channels: 1-4; extendable Valve oven: - independently heated, 40-150 °C - up to 3 valves - gas sampling or liquid sampling valves - diaphragm valves with internal purge Column ovens: - independently heated, 40-250 °C - up to 3 Detectors: TCD, FID, PDD, PED, PID, PFPD, MS |
Columns - capillary 0,32/0,25mm ID and smaller - wide bore MXT 0.53mm ID - micropacked 1/16” OD Column switching - diaphragm valve - Deans backflush or heartcut Sampling - inlet temperature: max. 150 °C, - inlet pressure: max. 21 bar (depending on used valve) - connection: 1/8” Swagelock - injection volume: 0.06 ul to 5ml (loop) Repeatability - 0.2% RSD or better (n=10, Ethane, TCD detection) Data communications - LAN for GC parameter control and data transfer (EZChrom/ChromQuest/OpenLab); GC editor available - 8 * external events 24V (valve control) - 4 * SSR output - 24 VDC - 3 * digital input (remote start-stop-prep) - optional analogue output (0-1V) for detector signals Environmental conditions - Ambient operating temperature: 5-45 °C - Ambient operating humidity: 5-95 %, (non-condensable) |