PAMAS S50 Applications:
Contamination control and Condition Monitoring of
- hydraulic oil systems
- lube oil systems
- test rigs
- parts cleaning
Pressure range and Sample flow:
Pressure: min. 0.2 to max. 20 bar
Flow: 5 to 50 ml/min, depending on system pressure
Viscosity:
Up to 1000 cSt, depending on system pressure
Volumetric particle sensor: PAMAS HX
Maximum particle concentration: 24.000 particles per ml at a coincidence quote of 7.8% and a flow rate of 25 ml/min
Calibration range: 4 µm(c) to 70 µm(c) according to calibration standard ISO 11171
Measurement in eight size channels: > 4 µm(c), > 6 µm(c), > 10 µm(c), > 14 µm(c), > 21 µm(c), > 25 µm(c), > 38 µm(c) and > 70 µm(c)
Report of measuring results according to cleanliness class standards SAE AS 4059 and ISO 4406. Particle counter measures ISO classes from 0 to 22.
Interfaces for data transfer
Standard equipment: RS485
Optional equipment: analogue 4-20 mA interface. Parallel data transmission for the size channels > 4 µm(c), > 6 µm(c), > 14 µm(c) and > 70 µm(c) or serial data transmission for all eight size channels.
Options:
PAMAS S50P: Online particle counter with integrated pump for flow control to a constant flow of 25 ml/min